INTRODUCTION

Methacrylate resins are one of the most commonly used materials in dentistry because of their biologic, physical, esthetic, and manipulative properties. Methacrylate polymers are capable of providing an excellent balance of features and characteristics needed for use in oral environment. Despite the various methods used to initiate the polymerization of denture base resins, the conversion of monomers to polymers is never complete; some unreacted (residual) monomer is always present.1-3 It has been proved that high level of residual monomer decreases the strength of denture base resins.4 It is also documented that autopolymerized resins have higher percentage of residual monomer (2–5%) than in heat cured polymers (0.2–0.5%).5,6 Autopolymerizing resins are commonly used in procedures like fabrication of temporary record bases, splints or stents, provisional restorations, orthodontic appliances, and repair of prosthesis. The strength of this material has always been a matter of concern for their long-term use. Several curing methods have been recommended to improve the overall quality of acrylic resin restorations. The methods include bench curing, curing under a coat of petroleum jelly, in monomer saturated atmosphere, and under controlled pressure in air and/or Luke warm water.7

But the search for newer methods with ease of operations for improving the strength to obtain more durable prosthesis is still on. Hence, this study was planned to evaluate the synergistic effect of variation in time and temperature at constant pressure for enhancing the overall strength of the appliances or prosthesis made using two autopolymerizing resins for their better utilization in dentistry.

MATERIALS AND METHODS

Materials selected for the study were:

  • RR Cold Cure resin (DPI – Bombay Burmah Trading Corporation Ltd, 9-Wallace Street, Mumbai) – D1

  • RR-Rapid Repair Powder (Dentsply India Pvt. Ltd, Plot No. 9, Phase-I, Udyog Vihar, Gurgaon) – D2.

Figs 1A and B

Die used in the study

johsr-8-15-g001.tif

The materials were designated as D1 and D2.

Methodology

Test was done in accordance with American Dental Association (ADA) specification no. 12. A precise metal die was fabricated for making the acrylic resin samples to be used in the study (Figs 1A and B). Samples were prepared as under:

  • Ten samples of materials D1 and D2 were prepared as control group and curing was carried out in open air at room temperature.

  • Five samples for materials D1 and D2 were fabricated and subjected to altered conditions of time and temperature with constant pressure of 30 psi using custom-made pressure pot as per the protocol of the study.

The details of sample size and groups are described in Table 1.

A brass die with five rectangular slots measuring 64 mm length, 10 mm width, and thickness of 2.5 mm was custom-made for the fabrication of samples. To facilitate the easy removal of samples, the die was made into two halves, which could be reassembled to form one unit when positioned onto a metallic base with the help of two metallic projections (P) of 2.5 mm height and corresponding holes (H) on either side for proper orientation.

Table 1: Details of sample size and groups

GroupsTemperatureTime interval (in minutes)No. of samples of D1No. of samples of D2
ControlRoom temperatureAs required1010
Group 140°C555
1055
1555
Group 250°C555
1055
1555
Group 360°C555
1055
1555
Group 470°C555
1055
1555

Autopolymerizing resin was mixed for 10 seconds according to manufacturer’s instruction to obtain the desired consistency and poured into the die. The entire assembly was then kept submerged in water in the custom-made pressure pot (Fig. 2) having provision of automatic controls for temperature, time, and pressure. Curing was carried out at specific temperature and time as per the protocol of the study (Table 1), at a constant pressure of 30 psi. Postcuring, the samples were kept in distilled water for 48 hours before subjecting them for testing.

Measuring the Flexural Strength

The cured samples were then tested for flexural strength on universal testing machine using three-point bending test. For this purpose, a specially designed fixture of dimensions 50 mm length, 30 mm width, and height of 30 mm was used along with a customized “T”-shaped stress applicator rod to apply stress at the center of the specimen. The readings were recorded to calculate the flexural strength using the formula:

johsr-8-15-e001.tif

where FS denotes the flexural strength, p is the peak load applied, l is the length of samples between two horizontal points “R,” b is the specimen width, d is the specimen thickness.

Fig. 2

Custom-made pressure pot

johsr-8-15-g002.tif
Graph 1

The mean flexural strength of materials D1 and D2, cured at different time and temperature, at a constant pressure

johsr-8-15-g003.tif

RESULTS

Descriptive analysis of D1 and D2 materials at various temperatures and time intervals at constant pressure is listed in Table 2. There was a continuous increase in strength from 40° to 70°C (Graph 1). Significant difference in strength was observed at 15 minutes in 50°C group and at each time interval of 60°C and 70°C for both the materials. There was no significant difference in flexural strength of above materials at 40°C and the control group.

Table 3 shows ‘Student’s unpaired t-test’ for comparison between flexural strength of control groups of materials D1 and D2. No significant difference was noticed in strength for control group of both the materials. At 40°C, for different time intervals, the values obtained were nonsignificant, for strength of both materials.

At 50°C, highly significant values were noticed at 15 minutes time interval for D2 material, whereas the values remained significant at 5 and 10 minutes, suggesting increased strength of D2 material with increase in time interval at a constant pressure. The values of D2 material showed highly significant values when compared to D1 material at 60° and 70°C, with gradual increase in time, thus indicating the increase in strength of D2 material for each time interval.

Table 2: The descriptive statistics showing mean, standard deviation and standard error mean of specimen of D1 and D2 materials

D1D2
TimeMeannStd. deviationStd. error meanMeannStd. deviationStd. error mean
40°5 minutes61.1050.570.2560.7051.310.58
10 minutes61.3051.710.7660.9551.320.59
15 minutes61.5650.880.3961.4850.910.40
50°5 minutes64.0350.890.3966.1751.400.62
10 minutes64.3250.630.2866.3551.100.49
15 minutes64.0650.970.4367.2950.740.33
60°5 minutes71.6351.050.4776.4250.760.34
10 minutes72.3150.470.2176.7251.370.61
15 minutes73.6750.570.2580.5650.730.32
70°5 minutes82.0650.590.2685.2250.840.37
10 minutes83.2350.520.2386.8650.800.36
15 minutes84.8950.410.1889.1050.670.30
Control56.52100.510.1656.63100.580.18

Table 3: A detailed comparison between two materials (Student’s t paired test)

t-test for equality of means
95% confidence interval of the difference
GroupTime intervalst-valuedfp-value  Mean differenceStd. error difference  Lower  Upper
40°C5 minutes0.62280.551 NS, p > 0.05  0.400.64–1.08  1.88
10 minutes0.35580.732 NS, p > 0.05  0.340.96–1.88  2.57
15 minutes0.14080.892 NS, p > 0.05  0.080.56–1.23  1.39
50°C5 minutes2.8680.021 S, p < 0.05–2.130.74–3.85–0.41
10 minutes3.5680.007 S, p < 0.05–2.030.57–3.35–0.71
15 minutes5.9280.000 HS, p < 0.0001–3.230.54–4.49–1.97
60°C5 minutes8.2380.000 HS, p < 0.0001–4.780.58–6.12–3.44
10 minutes6.7780.000 HS, p < 0.0001–4.410.65–5.91–2.90
15 minutes16.4680.000 HS, p < 0.0001–6.880.41–7.85–5.92
70°C5 minutes6.8280.000 HS, p < 0.0001–3.160.46–4.23–2.09
10 minutes8.3880.000 HS, p < 0.0001–3.620.43–4.62–2.62
15 minutes11.8480.000 HS, p < 0.0001–4.200.35–5.02–3.38
Control group0.425180.67 NS, p > 0.05–0.100.24–0.62  0.41
NS: Nonsignificance

DISCUSSION

It is a well-known fact that polymerization of chemically activated resin remains incomplete due to the presence of unreacted monomer, hence, the strength of material is questionable. Moreover, its wide application in dentistry demands superior physical and mechanical properties. The decreased strength of these resins is attributed mainly to the presence of unreacted monomer which acts as plasticizer, resulting in reduced flexural strength.

Many attempts have been advocated to reduce the amount of residual monomer and increase the strength of these resins, like processing of the resin in water under pressure to reduce porosity and increase the strength.8,9 Few researchers have favored the effect of temperature, rather than pressure, for increased strength of resin.10,11 Also soaking resin in water helps to decrease the amount of residual monomer, which ultimately increases the strength of the resin.12 Though water temperature seems to hasten polymerization, the temperature has not been adequately defined and also the influence of time on the curing under specific pressure and temperature is not being clearly mentioned in the literature. Hence, this study was conducted to evaluate the effect of increased temperature, time, at a constant pressure.

In the present study, it was observed that there is gradual increase in strength with rise in temperature from 40°C to 70°C at a constant pressure. Highest strength of the resin was obtained with the samples cured at 70°C, 40°C, and 50°C showed not much difference in the flexural strength, but definitely had increased strength than the samples cured in open air. In each group the strength obtained at 15 minutes was more than that of 5 and 10 minutes. The results obtained in the study are comparable with the results obtained by various researchers which showed increased strength at higher temperatures.8,9,11

The increase in strength can be attributed to the more complete polymerization at higher temperatures as heat activates the chemical reaction and the number of free radicles is lowered. The curing carried out inside the pressure unit in the pressure environment also has an effect on reducing porosity from inside the resin, which in turn enhances the strength. Reduction in the amount of residual monomer also helps in increase of strength of the resin. And it also has the added advantage of making the resin more biocompatible by reducing the tissue irritation caused by the monomer. Hence, it can be recommended to cure the resin under water at 70°C at a constant pressure for minimum of 10 minutes to obtain the desired strength of the autopolymerizing resins.

Clinical Implication

The results of this study reveal the advantage of placement of autopolymerizing resins at high temperatures and a constant pressure so as to provide sufficient strength and long-term durability of the resins. Based on the results of this study, placement of resin in hot water at controlled pressure will definitely enhance the strength of the resin and hence, it should be followed routinely in practice to provide more acceptable prosthesis.

CONCLUSION

The result of this study could provide with the data that will help the clinician and laboratory personale for better manipulation and usage of autopolymerizing resins. Within the scope and limitations of the study, the following conclusions were made as per the findings:

  • Curing the autopolymerizing resin in hot water under pressure is an effective means of increasing the strength of the resin.

  • The resin samples cured at 70°C, for 15 minutes under pressure of 30 psi, produced excellent result with increased flexural strength as compared to the samples of control group.

  • Time interval of 10 minutes is sufficient to provide adequate strength to the autopolymerizing resin sample, when compared with samples of control group.

  • Among the group of two materials used in the study, D2 material has showed more strength than D1 at temperatures of 60°C and 70°C, but the strength is almost the same at room temperature in open air.

Processing of resin as per the method mentioned in the study can be employed routinely in laboratory protocol to improve the strength and durability of the prostheses fabricated with autopolymerizing resin.